Medical devices recalls
2,678 recalls since January 2025 from Health Canada, FDA, newest first. Last notice 19 Nov 2026.
2,678recalls on record
Page 11 of 54
- FDABoston Scientific Corporation recalls FARADRIVE STEERABLE SHEATH CLEAR and 3 moreFDABoston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage…
- FDAParagon 28, Inc. recalls P28 PHANTON INTRAMEDULLARY NAIL SYSTEMFDAParagon 28 received two (2) complaints were received involving breaches in the sterile barrier of the pouch packaging.
- FDAOlympus Corporation of the Americas recalls Olympus 4K UHD LCD MonitorFDAPotential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities.
- FDAICU Medical Inc. recalls Hotline Fluid Warming SetFDAICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate…
- FDAMedline Industries, LP recalls MEDLINE ANTERIOR HIP and 1 moreFDAMedline Industries, LP is issuing a recall for specific Medline kits containing B. Braun convenience kits that include Bupivacaine Hydrochloride in…
- FDAFerno-Washington Inc recalls Power F2 Powered Fastening SystemFDAProduct may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case…
- FDAMedline Industries, LP recalls MEDLINE medical procedure convenience kit labeled asFDAThe procedure kits contain ICU Medical, Inc. pain management kits containing Bupivacaine Hydrochloride in Dextrose Injection, USP, which were…
- FDAMedline Industries, LP recalls Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters…FDAAffected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material…
- FDAMicro Therapeutics, Inc. recalls Pipeline Flex Embolization Device with Shield TechnologyFDAIncorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants…
- FDAMedline Industries, LP recalls Medline medical convenience kits labeled asFDAMedline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low…
- FDAMedtronic Neurosurgery recalls Ventriculostomy KitFDAVentriculostomy devices, used during ventriculostomy procedures to gain access to the ventricular system, may exhibit endotoxin levels that exceed…
- FDAKENT IMAGING, INC. recalls SnapshotGLO Wound Imaging DeviceFDAThe software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully…
- Health CanadaSigna MR SystemsHealth CanadaPerformance
- Health CanadaPatient Tables of Philips Allura and Azurion systemsHealth CanadaPerformance
- Health CanadaSphere-9™ CatheterHealth CanadaPerformance
- Health CanadaSpacelabs Healthcare Xhibit® Central StationHealth CanadaPerformance
- Health CanadaHeartSine® Samaritan® PAD SAM 350P/SAM 360P/SAM 500P - Membrane TailHealth CanadaPerformance
- Health CanadaOverStitch™ Suture Cinch and Suture Cinch–LongHealth CanadaPerformance
- Health CanadaDCA Vantage® HbA1c (DX Claim)Health CanadaPerformance
- Health CanadaHeartSine® Samaritan® PAD SAM 350P/SAM 360P/SAM 500PHealth CanadaLabelling and packaging
- FDAOlympus Corporation of the Americas recalls Olympus Ellik Evacuator Includes Glass Body and 3 moreFDASome sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used…
- FDAIntuitive Surgical, Inc. recalls da Vinci SP Instrument Arm DrapeFDARobotic-assisted surgical system arm drapes may detach from the instrument arm due to an issue involving the drape rings located at the top of the…
- FDAAbiomed, Inc. recalls ABIOMED Impella CP Set and 8 moreFDAA potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These…
- FDAPapablic Inc recalls Papablic Replacement Brush HeadsFDAPotential for brush heads to break.
- FDAArgon Medical Devices, Inc recalls General Biospy Tray and 5 moreFDAKits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for…
- FDAOscor Inc. recalls Abiomed REFFDADue to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along…
- FDAICU Medical, Inc. recalls ICU Medical Plum Duo Infusion Pump and 1 moreFDAIn a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the…
- FDAMedtronic Perfusion Systems recalls Medtronic Bio-Medicus Adult Cannula Kit and 1 moreFDAMedtronic received two complaints of loose foreign material confirmed as containing trace levels of blood.
- FDATosoh Bioscience Inc recalls ST AIA-Pack TestosteroneFDAThe Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA)…
- FDAOrthofix U.S. LLC recalls Orthofix MINIRAIL and 1 moreFDAUnilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail…
- FDA3B Medical, Inc. recalls Brand NameFDAPackaging and labeling discrepancy involving product size identification and associated Unique Device Identifier (UDI/GTIN).
- FDAMaquet Cardiovascular, LLC recalls The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model V…FDAGetinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma…
- FDAGE Medical Systems, LLC recalls Optima XR200amx and 2 moreFDAGE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of…
- FDAMedline Industries, LP recalls Medline Kits containing Zimmer Biomet mixing bowls and spatulas TOT… and 1 moreFDAKits contain Zimmer Biomet mixing bowls and spatulas, which were recalled due to seal nonconformities which present risk of sterility breach. Health…
- FDADaavlin Distributing Company recalls Photera 600 3-D and 2 moreFDACalibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were…
- FDABoston Scientific Corporation recalls Rapid Refill Continuous Injection SystemFDADevices contain recalled Medline Namic Angiographic Control Syringe with Rotating Adaptor. There is a risk of syringe loosening or disconnecting…
- Health CanadaMequ A/S ºM Warmer LeakageHealth CanadaPerformance
- Health CanadaShimadzu X-ray High Voltage GeneratorsHealth CanadaPerformance
- Health CanadaCardinal Health - Presource® Kits Containing the Ritmed® Neurological SpongesHealth CanadaSterility
- Health CanadaPhilips Achieva and Ingenia SystemsHealth CanadaPerformance
- Health Canada4K UHD LCD MonitorHealth CanadaPerformance
- Health CanadaMerge PACSHealth CanadaPerformance
- Health CanadaER-Spec Single-Use Vaginal Speculum with LED Light SourceHealth CanadaPerformance
- Health CanadaAtellica® CH Urine Albumin (UAlb)Health CanadaPerformance
- Health CanadaSolar™ GIHealth CanadaPerformance
- Health CanadaCardiac WorkstationHealth CanadaPerformance
- Health CanadaPhilips Allura Xper FD seriesHealth CanadaPerformance
- Health CanadaStrykeFlow PS and AHTOHealth CanadaPerformance
- Health CanadaSEMA 3Health CanadaDevice compatibility
- Health CanadaV-Loc™ 90 and 180 Suture ProductsHealth CanadaSterility