GE Medical Systems, LLC recalls Senographe Pristina and 1 more

Published
11 Nov 2026
Units affected
3 products
Agency
FDA
Country
United States

The hazard

GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Senographe Pristina

Senographe Pristina

Senographe Pristina 3D

Brand / company
GE Medical Systems, LLC
Codes / lots
GTIN(s): 00840682118460, 00840682142052, 00840682145879, 00195278022745, 00195278276971, 00195278661999, 00195278665379, 00195278914026, 00198953109281, 10198953054311, 10195278901764. | GTIN(s): 00840682118460, 00840682142052, 00840682145879, 00195278022745, 00195278276971, 00195278661999, 00195278665379, 00195278914026, 00198953109281, 10198953054311, 10195278901764.
Distribution
USA and OUS.
Recall status
Ongoing