Aerin Medical, Inc. recalls Aerin Medical VivAer Stylus

Published
15 Jan 2025
Units affected
479 units
Agency
FDA
Country
United States

The hazard

Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.

Brand / company
Aerin Medical, Inc.
Codes / lots
Lot #10235, exp. 2026-05-01; UDI (01)08886479300269(17)260501(10)10235
Distribution
US Nationwide Distribution: AL, AZ, CA, CO, GA, ID, IL, IN, KY LA, MI, MN, NY, NE, NY, OH, OR PA, SC, TN, TX, UT, VA, WA, WI, and WV.