Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver

Published
15 Jan 2025
Units affected
2265
Agency
FDA
Country
United States

The hazard

Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.

Brand / company
Spacelabs Healthcare, Inc.
Codes / lots
Service Manuals P/N 070-2409-07 and Prior. UDI-DI: 10841522107177, 10841522100246
Distribution
US: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX, ID, NJ, FL, OH, MS, OR, AK, IL, AR, KS, AL, TN, WI, VA, NM, CA, IN, RI, NV, MA, NC, ME, CT, AZ, MD, PR. OUS: Kuwait, Australia, Canada, Roumanie, Greece, Philippines, Spain, Italy, Hong Kong, Switzerland, Panama, Taiwan, Netherlands, Malaysia, Poland, Mexico, United Kingdom, Saudi Arabia, France, Czech Republic
Recall status
Ongoing