CheckMyRecalls

Mercury Enterprises, Inc. dba Mercury Medical recalls Neo-Tee T-Piece…

Mercury Enterprises, Inc. dba Mercury Medical recalls Neo-Tee T-Piece Resuscitator

Published
22 Jan 2025
Units affected
37370
Agency
FDA
Country
United States

The hazard

Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, 1050808, 1050809, 1050810, 1050811, 1050814, 1050832, 1050839, 1050840, 1050841, 1050842

Brand / company
Mercury Enterprises, Inc. dba Mercury Medical
Codes / lots
Part/UDI-DI/Lot: 1050805/10641043508053, 30641043508057/2421450805, 2423550805, 2423650805, 2426750805; 1050808/10641043508084, 30641043508088/2426450808, 2426750808, 2426850808, 2429050808, 2429350808; 1050809/10641043508091, 30641043508095/2426450809, 2426750809, 2429050809, 2429150809, 2429350809, 2429750809, 2429850809, 2430050809, 2430250809; 1050810 /10641043508107, 30641043508101/2426750810, 2429350810, 2429750810, 2429850810, 2430250810; 1050811/10641043508114, 30641043508118/2426550811, 2426650811, 2426750811, 2429050811, 2429150811, 2429250811, 2429350811, 2429450811, 2429750811, 2429850811, 2429950811, 2430250811; 1050814/10641043508145, 30641043508149/2426450814, 2426750814, 2429050814, 2429750814, 2430250814; 1050832/10641043508329, 30641043508323/2429050832, 242985083; 1050839/10641043508398, 30641043508392/2430350839; 1050840/10641043508404, 30641043508408/2429050840, 2429750840; 1050841/10641043508411, 30641043508415/2429050841; 1050842/10641043508428, 30641043508422/2426750842
Distribution
Worldwide Distribution: US (nationwide) to states of: PA, FL, IN, WI, ME, SC, WY, ID, MT, TX, MI, LA, MS, AR, WV, MD, KY, NJ, WA, CA, GA, MO, MA, NC, NV, TN, VA, OK, ND, MN, AL, HI, UT, NY, AZ, IL, OH, IA, NE, CT, AK, OR and OUS (foreign) to countries of: Hungary, Switzerland, Hong Kong, Spain, Canada, Italy, United Kingdom, United Arab Emirates, Saudi Arabia, France, Netherlands
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 22 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.