The hazard

Potential failure to alarm for "No Device Data" INOP. The device to alarm after a Hamilton ventilator is disconnected from the EC10 Module or IntelliBridge I/O board.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

IntelliVue Patient Monitor MX400. Model Number: 866060

IntelliVue Patient Monitor MX450. Model Number: 866062

IntelliVue Patient Monitor MX550. Model Number: 866066

IntelliVue Patient Monitor MX500. Model Number: 866064

IntelliBridge EC10 Module. Model Number: 865115

Brand / company
Philips North America Llc
Codes / lots
Model Number: 866060. UDI-DI Number: (01)00884838038752(21). Serial Numbers: DE35174105, DE35174084, DE35174083, DE35174082, DE35174080, DE35107369, DE35108320, DE35160410, DE671M4116, DE7770DA9W, DE35114616, DE35114618, DE671E8921, DE671E8917, DE7580AH50, DE7580AHEH, DE35123130, DE671V9297, DE671V9296, DE671V9298, DE35160989, DE35160994, DE35160988, DE35161272, DE35161273, DE35161274, DE7580B537, DE7580B52H. | Model Number: 866062. UDI-DI: (01)00884838038769(21); Serial Numbers: DE35169761, DE671V3903, DE671V3885, DE671V7898, DE7770DC1L. UDI-DI: 884838045859; Serial Numbers: DE32303399, DE32302479, DE32302480, DE32306956, DE35131294, DE35114613, DE35114612, DE671T9641, DE7770C86L, DE7770C86K, DE7770C86J, DE7770C86H, DE7770C86G, DE35120431, DE671K3731, DE671K3729, DE671K3706, DE671K3796, DE671K3816, DE671K3737, DE671K3773, DE671L0436, DE671L0462, DE671L0406, DE671L5214, DE671L5207, DE671L5206, DE671L5181, DE671L5169, DE671M3993, DE671M4048, DE671M4036, DE671M4037, DE671M4053, DE671M7009, DE671M7012, DE671M9968, DE671M9972, DE671M9963, DE671N0009, DE671N0010, DE671N0092, DE671N0008, DE671N0001, DE671N0002, DE671M9985, DE671M9999, DE671N0000, DE671P0411, DE671P1094, DE671P2457, DE671
Distribution
US distribution to Florida, North Carolina, and Oregan. International distribution to Canada, Estonia, France, Germany, Japan, Netherlands, Spain, Switzerland, United Kingdom.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 12 Feb 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.