Telcare, LLC recalls Philips Connected Blood Glucose Meter
- Published
- 12 Mar 2025
- Units affected
- 34,764 units
- Agency
- FDA
- Country
- United States
The hazard
Telcare, LLC has received customer complaints related to battery expansion/swelling for its Philips Connected Blood Glucose Meter (i.e., BGM 4 brand name) Monitoring System.. The issue was identified by post market surveillance data (customer complaints) - events of battery expansion/swelling and associated symptoms, e.g., leading to the device's removable back cover (battery port) to separate. If there is a loss of primary function of the device due to the described device defect, an immediate health consequence that may result is prolonged hypoglycemia or hyperglycemia due to delay in treatment, as the blood glucose measurement is not available to guide treatment. In a worst-case situation, there is potential for battery swelling to result in explosion of the battery and/ or device. Exposure to explosion, and subsequently projectiles from the explosion, may result in several immediate health consequences, including concussion, headache, dizziness, eye irritation, blurred vision, eye injury, and vision impairment.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Philips Connected Blood Glucose Meter, BGM 4; Model Number: TM0009;
- Brand / company
- Telcare, LLC
- Codes / lots
- Model Number: TM0009; UDI-DI: 00859519002407; Serial Numbers: Serial Numbers are formatted as TFXXXXXX or TGXXXXXX (X represents digits). Meter Serial Numbers: TF048802, TF098191, TF079123, TF076786, TF076836, TF089080, TF088984, TF086097, TF085390, TF085029, TF084238, TF098751, TF098975, TF098976, TF096704, TF096961, TF096966, TF117862, TF127077, TF049583, TF047153, TF063135, TF127267, TF100376, TF086287, TF104396, TF104399, TF104401, TF104403, TF104404, TF104405, TF104407, TF104408, TF104409, TF104410, TF104411, TF104412, TF104414, TF104415, TF104416, TF104417, TF104418, TF104421, TF104426, TF104427, TF104428, TF104429, TF104430, TF104431, TF104432, TF104433, TF104439, TF104440, TF104441, TF104442, TF104444, TF104446, TF104448, TF104449, TF104452, TF104453, TF104454, TF104455, TF104456, TF104458, TF104460, TF104461, TF104462, TF104467, TF104468, TF104469, TF104471, TF104473, TF104474, TF104479, TF104480, TF104481, TF104482, TF104483, TF104486, TF104487, TF104490, TF104491, TF104492, TF104494, TF104496, TF104498, TF104500, TF104501, TF104503, TF104505, TF104506, TF104510, TF104511, TF104512, TF104514, TF104516, TF104520, TF104521, TF104522, TF104525, TF104526, TF104527, TF104528,
- Distribution
- US Nationwide distribution.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 12 Mar 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.