CheckMyRecalls

DeRoyal Industries Inc recalls DeRoyal REF 71-C5100NS Tubing

Published
12 Mar 2025
Units affected
714 cases
Agency
FDA
Country
United States

The hazard

Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement

Brand / company
DeRoyal Industries Inc
Codes / lots
71-C5100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578738 Lot Number: CNWKD04-04 71-C6100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578851 Lot Numbers: CNWKD04-04 CNWKD04-05 CNWKD04-10 CNWKD09-08 71-C7100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578967 Lot Numbers: CNWKD04-06 CNWKD04-07
Distribution
U.S.: IL, MN, NC, PA, and VA O.U.S.: N/A
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 12 Mar 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.