INTELERAD MEDICAL SYSTEMS INCORPORATED recalls IntelePACS (Image…

INTELERAD MEDICAL SYSTEMS INCORPORATED recalls IntelePACS (Image Fusion Module)

Published
5 Mar 2025
Units affected
88
Agency
FDA
Country
United States

The hazard

A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

IntelePACS (Image Fusion Module) - InteleViewer

Brand / company
INTELERAD MEDICAL SYSTEMS INCORPORATED
Codes / lots
UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later
Distribution
Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia
Recall status
Ongoing