INTELERAD MEDICAL SYSTEMS INCORPORATED recalls IntelePACS (Image…
INTELERAD MEDICAL SYSTEMS INCORPORATED recalls IntelePACS (Image Fusion Module)
- Published
- 5 Mar 2025
- Units affected
- 88
- Agency
- FDA
- Country
- United States
The hazard
A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
IntelePACS (Image Fusion Module) - InteleViewer
- Brand / company
- INTELERAD MEDICAL SYSTEMS INCORPORATED
- Codes / lots
- UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 5 Mar 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.