TAS Medical Inc recalls TISSUE APPROXIMATION SYSTEM (TAS)
- Published
- 12 Mar 2025
- Units affected
- 114
- Agency
- FDA
- Country
- United States
The hazard
Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP
- Brand / company
- TAS Medical Inc
- Codes / lots
- Model/UDI-DI/Lot(Expiration): T-4000/00197644217267/230403(2024/04/20); T-5000/00860011435921/240801(2025/08/28), 240901(2025/09/25); T-LAP/00860011435914, 00197644927241/240702(2025/07/08), 230906(2024/11/29), 230905(2024/11/29)
- Distribution
- US Nationwide distribution in the states of LA, OK, TN.
- Recall status
- Ongoing