GE Healthcare (China) Co., Ltd. recalls Brivo MR355 and 4 more
- Published
- 5 Mar 2025
- Units affected
- 5 products
- Agency
- FDA
- Country
- United States
The hazard
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Brivo MR355, NMRI system
SIGNA Creator, SIGNA Explorer, NMRI systems
SIGNA MR355, SIGNA MR360, NMRI system
SIGNA MR380, NMRI system, Non-US Only
Optima MR360, NMRI system
- Brand / company
- GE Healthcare (China) Co., Ltd.
- Codes / lots
- Brivo MR355 (No UDI) System IDs 082427010096 082427020046 082427040126 082427070076 082427070077 082427070170 082427100245 082427120151 082427180030 082427210147 082427210148 082427220025 082427220045 082427230060 082427240060 082427310104 0856270058 0856270059 MRR10200 MRR9920 RU1063MR03 RU8902MR01 (Updated 12/29/25 - Additional System ID 83027805544513) (Updated 1/28/26 - Additional System IDs: 82427220047 82427240055 82427070163 82427060068 82427140137 82427010122 82427280018) (Updated 2/19/26: Additional System ID EG1736MR02) | 1. Creator A. UDI-DI 00195278577221, System ID MR00160 MR00171 MR00189 THMR54503051 (Update 12/29/25 - New System ID MRR10626) (Update 1/28/26 - New System IDs MRR10269 850270138) (Update 2/19/26 - Addl System IDs MR00299¿ PHMRI24836517 12440397 13131449¿ PC0084) B. UDI-DI 00195278577238, System ID 083027026539324 5973MR0087 5973MR0091 5973MR0095 5973MR0096 5973MR0097 5973MR0100 5973MR0103 5973MR0104 5973MR0106 5973MR0107 5973MR0109 9830MR0112 9830MR0113 9830MR0114 9830MR0117 A199750201 IDMRI24467771 MR00082 MR00111 MR00120 MR00138 MR00140 MR00143 MR00152 MR00157 MR00174 MR00179 MR00180 MR00187 MR00191 MR00192 MR00195 MR00210 MR00229 MR00231 MR00236 MR00
- Distribution
- Domestic US distribution nationwide. International distribution worldwide.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 5 Mar 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.