CheckMyRecalls

Merge Healthcare, Inc. recalls VERICIS and 1 more

Published
26 Mar 2025
Units affected
5 products
Agency
FDA
Country
United States

The hazard

When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

VERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.1, 12.1.1, 12.2, 12.3, 12.3.1, 12.3.2, 12.3.3, 12.4, 12.4.1, 12.4.2; Merge Cardio is a software medical device that is an integrated multi-modality image cardiovascular information management system.

VERICIS, Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.0.1, 12.0.2; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

VERICIS, Merge Cardio, Model/Catalog Number: Version 11.x, Software Version: 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

Brand / company
Merge Healthcare, Inc.
Codes / lots
The following versions were introduced prior to the implementation of UDI. Version Numbers: 9.0.6, 9.0.8, Part Numbers: 88-00061-00, 88-00062-00, 88-00063-00, 88-00054-00, 88-00064-00, | Software Versions and Part Numbers: 10.0: Part Numbers: 88-00527-00, 88-00528-00, 88-00529-00, 88-00530-00, 88-00531-00, 88-00532-00, 88-00533-00, 70-00008-00; 10.1.1.1: Part Numbers: 88-01094-00; 10.1.2: Part Numbers: 88-01074-00, 88-01075-00, 88-01076-00, 88-01077-00, 88-01078-00, 88-01079-00, 88-00781-00, 88-00803-00, 88-01080-00, 70-00473-00; 10.2.0: Part Numbers: 88-01115-00, 88-01116-00, 88-01117-00, 88-01118-00, 88-01119-00, 88-01120-00, 88-01121-00, 88-01122-00, 88-01123-00, 70-00491-00, 10.2 P2: Part Numbers: 88-01203-00 10.3 P1 UDI - (01)00842000100041(10)10.3.0.2009(11)170309 | Lot Code: Software versions and UDIs 12.4.1 UDI: (01)00842000101123(10)12.4.1(11)240522 12.4.2 UDI: (01)00842000101123(10)12.4.2(11)250206 12.1, UDI: (01)00842000100928(10)12.1(11)210122; 12.1.1, UDI: (01)00842000100928(10)12.1.1(11)211220; 12.2 UDI: (01)00842000100065(10)12.2.0(11)220713; 12.3, UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.1 UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.2 UDI: (01)0084200
Distribution
US
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 26 Mar 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.