Bard Access Systems, Inc. recalls BD 4 Fr Single-Lumen PowerPICC…
Bard Access Systems, Inc. recalls BD 4 Fr Single-Lumen PowerPICC Catheters
- Published
- 23 Apr 2025
- Units affected
- 718,461
- Agency
- FDA
- Country
- United States
The hazard
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
BD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155, CK000540, 3174118, CK000483, 3174335, 3174355, CK000196, 9174118, CK000955, 3174135, CK000617, CK000839, 3194355, 6174118, 3194135, 3194335, 2174108, 9194118, 3194118, 1174118, CK000285C, CK000335A, CK000796, 1194108D, CK000757, CK001042, 9174108, 1174108D1, CK000631B, 3194155, 3174108, CK000567, 1194108D4, 1174108D2, CK000677, CK000864, CK000937A, CK000930, CK001027, 1174108D, CK000867, CK000900, CK000315A, CK000589A, CK000854B, CK000897, CK000953A, CK000586B, CK000549, CK000844, 1194108D1, 1174108D4, 1194108D3, CK000934, CK000986, CK000411, CK001018, 1174108, CK000060, CK000513A, CK001021, CK001030, CK000501A, CK000619A, CK000903, 3194108, 1194108D2, 9194108D, CK000959, CK000613, CK000991, CK000983, CK000881, CK000721, CK000642, CK000289C, CK000894A, CK000914, CK000445, CK000687, CK000524A, CK000819, CK000266A, 1174108D3, 3194108D, CK000545, 1174108D5, 3174108D, CK000775A, 1194108D5, CK000736, CK000739, 9194108, 1194108, CK000660, CK000609A, 9174108D, CK001067, CK000077D, CK000882, CK000874, CK000994, CK000870A, CK000443C, CK000668B, CK000784
- Brand / company
- Bard Access Systems, Inc.
- Codes / lots
- REF/UDI-DI/Lot(Expiration): 3174155/00801741027611/REGQ0844(6/30/2025), REGZ1439(11/30/2025), REHP0860(11/30/2025), REHQ3097(6/30/2026), REHR0022(6/30/2026), REHT1151(7/31/2026), REHU1525(11/30/2026), REHU2455(6/30/2026), REHU2614(11/30/2026), REHV2765 (7/31/2026), REJN3472(2/28/2027), REJP4216(5/31/2027), REJQ2272(7/31/2027), REJR1411(8/31/2027), REJR2565(4/30/2027), REJR2614(8/31/2027), REJT0755(9/30/2027); CK000540/00801741105142/REGZ1425(10/31/2025), REHP0808(11/30/2025), REHU1513(1/31/2026), REHV2221(1/31/2026), REHY2842(10/31/2026), REJN4145 (12/31/2026), REJR1322(3/31/2027), REJS2110(4/30/2027), REJS3773(2/28/2027); 3174118/00801741027598/REGZ2512(11/30/2025), REHP1760(1/31/2026), REHP2603(1/31/2026), REJN1059(12/31/2026); CK000483/00801741099397/REHN3187(10/31/2025), REHQ1804(6/30/2026), REJN1065(5/31/2027), REJP2018(6/30/2027), REJR1321(8/31/2027); 3174335/00801741027628/REHN3908(10/31/2025), REHP0800(10/31/2025), REHQ3070(11/30/2025), REHT1129(5/31/2026), REHU1519(6/30/2026), REHV1391(7/31/2026), REHV3050(7/31/2026), REHZ1011(11/30/2026), REJN4109(12/31/2026), REJQ0405(2/28/2027), REJR2090(3/31/2027), REJT0833(5/31/2027); 3174355/00801741027635/REHQ1781(11/30/2025), REHS1
- Distribution
- Worldwide - US Nationwide distribution including in the states of WI, NC, TX, NY, CA, MN, MT, RI, MA, OH, CT, SC, IL, CO, VA, NJ, GA, UT, FL, PA, IN, MD, KY, AZ, MI, WA, HI, MO, VT, AL, NM, TN, DE, LA, IA, NH, ID, DC, WV, KS, NE, SD, ND, WY, AK, OK, OR, AR and the countries of Andorra, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, French Polynesia, French Guayana, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Luxembourg, Martinique, Mexico, Netherlands, New Caledonia, New Zealand, Norway, Pakistan, Poland, Portugal, Qatar, Republic of Korea, Reunion, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 23 Apr 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.