Zimmer, Inc. recalls Zimmer Biomet Ceramic Head
- Published
- 30 Apr 2025
- Units affected
- 5 products
- Agency
- FDA
- Country
- United States
The hazard
The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +0 Neck, Model Number 802802202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +3 Neck, Model Number 802602203; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +3 Neck, Model Number 802802203; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, -3 Neck, Model Number 802802201; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented
- Brand / company
- Zimmer, Inc.
- Codes / lots
- GTIN 00887868376351, All lots distributed with IFU 87-6204-051-99 Rev D or prior | GTIN 00887868376337, All lots distributed with IFU 87-6204-051-99 Rev D or prior | GTIN 00887868376320, All lots distributed with IFU 87-6204-051-99 Rev D or prior | GTIN 00887868376368, All lots distributed with IFU 87-6204-051-99 Rev D or prior | GTIN 00887868376344, All lots distributed with IFU 87-6204-051-99 Rev D or prior
- Distribution
- Worldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 30 Apr 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.