Abbott Molecular, Inc. recalls The Alinity m System is designed to…
Abbott Molecular, Inc. recalls The Alinity m System is designed to automate the steps for nucleic …
- Published
- 30 Apr 2025
- Units affected
- N/A
- Agency
- FDA
- Country
- United States
The hazard
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).
- Brand / company
- Abbott Molecular, Inc.
- Codes / lots
- All Serial Numbers. 08N53-001, UDI-DI: 00884999047389; 08N53-002, UDI-DI: 00884999048034; 08N53-032, UDI-DI: 00884999047587
- Distribution
- Worldwide - US Nationwide distribution.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 30 Apr 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.