Globus Medical, Inc. recalls MAGEC 2 Rod
- Published
- 30 Apr 2025
- Units affected
- 1 unit
- Agency
- FDA
- Country
- United States
The hazard
Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;
- Brand / company
- Globus Medical, Inc.
- Codes / lots
- Version/Model: MC2-5090S; UDI-DI (with lot nuber): (01)00887517834041(11)250311(17)300311(10)GB1607ET
- Distribution
- US Nationwide distribution in the state of AR.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 30 Apr 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.