Siemens Healthcare Diagnostics Inc recalls Stratus CS Acute Care cTNI…
Siemens Healthcare Diagnostics Inc recalls Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I …
- Published
- 7 May 2025
- Units affected
- 367 units
- Agency
- FDA
- Country
- United States
The hazard
Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for cardiac troponin I in heparinized plasma by removal. The reason for the recall is an increased occurrence of random non-repeatable false positive cardiac Troponin l(cTnI) results at any point during the TestPaks shelf life when using the Stratus CS cTnI Acute Care Testpak.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.
- Brand / company
- Siemens Healthcare Diagnostics Inc
- Codes / lots
- Siemens Materials No. 10445071; UDI-DI: 0405686902174VK; Lot No. 234337002 and forward.
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, IL, KS, MI, MO, NJ, NY, OH, OK, PA, SC & WI. The countries of Austria, Canada, Chile, France, Germany, Iraq, Italy, Malaysia, North Macedonia, Saudi Arabia, Serbia, Slovenia, Spain, Sweden, Switzerland & Vietnam.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 7 May 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.