GE Medical Systems China Co., Ltd. recalls GE Healthcare CARESTATION…
GE Medical Systems China Co., Ltd. recalls GE Healthcare CARESTATION 620 A2 and 14 more
- Published
- 7 May 2025
- Units affected
- 15 products
- Agency
- FDA
- Country
- United States
The hazard
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
GE Healthcare CARESTATION 620 A2, Model/REF Number 1012-9620-002. Anesthesia system.
GE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system.
GE Healthcare CARESTATION 750 A2, Model/REF Number 1012-9750-002. Anesthesia system.
GE Healthcare Carestation 650 SE, Model/REF Number 1012-9650-012. Anesthesia system.
GE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200. Anesthesia system.
GE Healthcare CARESTATION 650 A2, Model/REF Number 1012-9650-002. Anesthesia system.
GE Healthcare CARESTATION 650C A2, Model/REF Number 1012-9655-002. Anesthesia system.
GE Healthcare Carestation 620 SE, Model/REF Number 1012-9620-012. Anesthesia system.
GE Healthcare CARESTATION 750c A2, Model/REF Number 1012-9755-002. Anesthesia system.
GE Healthcare CARESTATION 650 A1, Model/REF Number 1012-9650-000. Anesthesia system.
GE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200. Anesthesia system.
GE Healthcare CARESTATION 750 A1, Model/REF Number 1012-9750-000. Anesthesia system.
- Brand / company
- GE Medical Systems China Co., Ltd.
- Codes / lots
- GTIN 00840682124546, Serial Numbers: SM623510024WB, SM623510028WB, SM623510029WB, SM623510030WB, SM623510031WB, SM623510032WB, SM623510025WB, SM623510026WB, SM623510027WB, SM624390041WB, SM624390042WB, SM623510033WB, SM623510034WB, SM624220006WB, SM624220008WB, SM623160001WB, SM623160002WB, SM624120009WB, SM624120010WB, SM624510014WB, SM624510015WB, SM624510016WB, SM624250055WB, SM624250056WB, SM624120013WB, SM624120015WB, SM624320001WB, SM623250007WB, SM623250011WB, SM624250057WB, SM624250058WB, SM624250059WB, SM624250060WB, SM624250061WB, SM624320002WB, SM624370013WB, SM624370014WB, SM624370015WB, SM624370016WB, SM624370017WB, SM624370018WB, SM623460225WB, SM623460226WB, SM623460227WB, SM623460228WB, SM623460229WB, SM623460230WB, SM623460231WB, SM624120007WB, SM624120008WB, SM623250005WB, SM623250006WB, SM623460223WB, SM623460224WB, SM623390147WB, SM624120011WB, SM624120012WB, SM624120014WB, SM623250001WB, SM623250002WB, SM623250003WB, SM623250004WB, SM623390148WB, SM623390149WB, SM623390150WB, SM623390151WB, SM623390152WB, SM624120001WB, SM624120002WB, SM624120003WB, SM624120004WB, SM624120005WB, SM624120006WB, SM624120016WB, SM624120017WB, SM624120018WB, SM623140001WB, SM623140
- Distribution
- Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T¿rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 7 May 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.