ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM)
- Published
- 21 May 2025
- Units affected
- 12 US, 3 OUS
- Agency
- FDA
- Country
- United States
The hazard
Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2
- Brand / company
- ICU Medical, Inc.
- Codes / lots
- LifeShield Infusion Safety Software Suite v2.1 and v2.2 UDI-DI 00887709122802
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, FL, IA, KY, LA, MA, MS, NY, PA, TN, and VA. The country of Philippines.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 21 May 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.