Covidien recalls Newport HT70 Ventilator
- Published
- 18 Jun 2025
- Units affected
- 5394
- Agency
- FDA
- Country
- United States
The hazard
Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-EU, HT70M-WW-NA, HT70-SY-EU. Newport HT70Plus Ventilator, REF: HT70PM-ES-EU, HT70PM-ES-NA, HT70PM-JP-NA, HT70PM-PT-BR, HT70PM-SY-AS, DLHT70PM-WW-NA, HT70PM-SY-NA, HT70PM-SY-UK, HT70PM-WW-EU, HT70PM-WW-UK, HT70PM-SY-EU, HT70PM-WW-NA. HT70-2 CONTROL BOARD ROHS, REF: GR-PCB3252A. HT70-2 CONTROL BOARD ROHS X1, REF: GR105796. HT70P CONTROL BOARD ROHS X1, REF: GR105795. HT70 Conversion Kit, REF: GRFSKHT70M-2.
- Brand / company
- Covidien
- Codes / lots
- REF/UDI-DI/Serial Numbers: HT70M-JP-NA/00839935008221/N10HT700710132, N10HT700710135, N10HT700710136, N10HT700710159, N10HT700710161, N10HT700710162, N10HT700710163, N10HT700710164, N10HT700810194, N10HT700810199, N10HT700810202, N10HT700810204, N10HT700810205, N10HT700810206, N10HT700910223, N10HT700910227, N10HT701010279, N10HT701011015, N10HT701011016, N10HT701111044, N10HT701111048, N10HT701111049, N10HT701111050, N11HT700111137, N11HT700111139, N11HT700111141, N11HT700111142, N11HT700111143, N11HT700111144, N11HT700211145, N11HT700211155, N11HT700211160, N11HT700211161, N11HT700211163, N11HT700211164, N11HT700311168, N11HT700311169, N11HT700311170, N11HT700311183, N11HT700311184, N11HT700311185, N11HT700311186, N11HT700311188, N11HT700311189, N11HT700311191, N11HT700311192, N11HT700311193, N11HT700311195, N11HT700311196, N11HT700311197, N11HT700311201, N11HT700311202, N11HT700311203, N11HT700311204, N11HT700311207, N11HT700311208, N11HT700311209, N11HT700311233, N11HT700311234, N11HT700311243, N11HT700311244, N11HT700311245, N11HT700311246, N11HT700311247, N11HT700311248, N11HT700311250, N11HT700311251, N11HT700311253, N11HT700311255, N11HT700411256, N11HT700411257, N11HT70041
- Distribution
- Worldwide - US Nationwide distribution including in the states of MA, CT, KS, OH, PR, NH, GA, WI, MS, VA, CO, TX, MN, AL, IA, NY, UT, NM, KY, RI, SD, PA, IA, CA, MO, HI, MI, IL, MT, NJ, TN, SC, FL, IN, AZ, ID, LA, WA, NC, MD, WY, NE, NV, ND, ME, AK, DE, OK, OR and the countries of Japan, Canada, Singapore, Lithuania, Kazakhstan, Slovenia, Colombia, China, Qatar, United Arab Emirates, Greece, Saudi Arabia, Nigeria, Romania, Netherlands, Lebanon, Oman, Jordan, Malawi, India, Croatia, Portugal, Turkey, Brazil, Pakistan, Trinidad and Tobago, Slovakia, Chile, Egypt, Switzerland, South Africa, Thailand, Bangladesh, Myanmar, Peru, Uruguay, Hungary, Argentina, Iran, Islamic Republic Of, Costa Rica, Vietnam, Australia, Guatemala, Mexico, Guyana, Jamaica, Dominican Republic, Bahamas, Bosnia and Herzegovina, South Korea, Tunisia, Ukraine, Uzbekistan, Indonesia, Burkina Faso, Italy, Panama, Sri Lanka, Malaysia, Namibia, Swaziland, Spain, Philippines, Bhutan, Mauritius, Poland, France, Cambodia, Czechia, Israel, Norway, Russian Federation, United Kingdom, Canary Islands, Kenya, Czech Republic, Croatia , Estonia, Maldova, Palestine.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 18 Jun 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.