SunMed Holdings, LLC recalls Adult Manual Resuscitator with Medium…
SunMed Holdings, LLC recalls Adult Manual Resuscitator with Medium Adult Mask
- Published
- 11 Jun 2025
- Units affected
- 11,358
- Agency
- FDA
- Country
- United States
The hazard
Affected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there would be interruption or delay in patient resuscitation, which may lead to life threatening consequences, including hypoxia, hypercapnia, organ failure, and death.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F
- Brand / company
- SunMed Holdings, LLC
- Codes / lots
- UDI-DI: EA: 10884389164822 CS: 40884389164823 Lots: 526782 526790 526796 526797 526798 526800 526802 526804 526805 526806 526807 526808 526809 526810 526814 526815 526816 526811 526817 526818
- Distribution
- US Nationwide distribution via Medline.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 11 Jun 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.