VIRTUAL INCISION CORPORATION recalls MIRA SURGICAL SYSTEM
- Published
- 11 Jun 2025
- Units affected
- 3 units
- Agency
- FDA
- Country
- United States
The hazard
The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3
- Brand / company
- VIRTUAL INCISION CORPORATION
- Codes / lots
- UDI-DI (01)00850038042028; Lot Numbers: 1020040211, 1020040212, 1020040213
- Distribution
- US Distribution to States: CA, FL, NE
- Recall status
- Completed
Source: U.S. Food and Drug Administration, published 11 Jun 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.