PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion R1.x and R2.x…

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion R1.x and R2.x systems

Published
18 Jun 2025
Units affected
9,707 units (2300 US, 7407 OUS)
Agency
FDA
Country
United States

The hazard

A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Azurion R1.x and R2.x systems; Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (7) 722221, (8) 722222, (9) 722223, (10) 722224, (11) 722225, (12) 722226, (13) 722227, (14) 722228, (15) 722280, (16) 722281, (17) 722282; Software Version: All software versions;

Brand / company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Codes / lots
Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (7) 722221, (8) 722222, (9) 722223, (10) 722224, (11) 722225, (12) 722226, (13) 722227, (14) 722228, (15) 722280, (16) 722281, (17) 722282; UDI-DIs: (1) 00884838085275, (2) 00884838085282, (3) 00884838085350, (4) 00884838085367, (5) 00884838085251, (6) 00884838085268, (7) 00884838099203, (8) 00884838099210, (9) 00884838099241, (10) 00884838099258, (11) 00884838099265, (12) 00884838099272, (13) 00884838099227, (14) 00884838099234, (15) Not commercialized in the US, (16) Not commercialized in the US, (17) Not commercialized in the US;
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, IRAN, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.
Recall status
Ongoing