STRATASYS LTD recalls TrueDent Clear and 4 more
- Published
- 16 Jul 2025
- Units affected
- 5 products
- Agency
- FDA
- Country
- United States
The hazard
Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.
Classification: Class III — this product is unlikely to cause health problems but breaks FDA rules (for example labeling).
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
TrueDent Clear, TDM 100, Model No. OBJ-09164 Light-curable methacrylate-based resin that enables fabrication of dental appliances
TrueDent Yellow, TDM140, Model No. OBJ-09169 Light-curable methacrylate-based resin that enables fabrication of dental appliances
TrueDent Cyan, TDM 120, Model No. OBJ-09165 Light-curable methacrylate-based resin that enables fabrication of dental appliances
TrueDent Magenta, TDM130, Model No. OBJ-09166 Light-curable methacrylate-based resin that enables fabrication of dental appliances
TrueDent White, TDM110, Model No. OBJ-09168 Light-curable methacrylate-based resin that enables fabrication of dental appliances
- Brand / company
- STRATASYS LTD
- Codes / lots
- UDI-DI 7290018915004 Lots 22994 (exp. 02-Apr-2024) and 23425 (exp. 07-Nov-2024) | UDI-DI 7290018915806 Lot 22861 (exp. 01-Feb-2024) | UDI-DI 7290018915400 Lots 23485 (exp. 08-Nov-2024) and 22904 (exp. 09-Feb-2024) | UDI-DI 7290018915608 Lot 23123 (exp. 01-May-2024) | UDI-DI 7290018915202 Lot 22377 (exp. 12-Oct-2023)
- Distribution
- US Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Maryland, Massachusetts, Michigan, Nebraska, New York, North Carolina, North Dakota, Oklahoma, Oregon, Utah, Wisconsin.
- Recall status
- Completed
Source: U.S. Food and Drug Administration, published 16 Jul 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.