Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Systems…
Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack)
- Published
- 6 Aug 2025
- Units affected
- 1,477 units
- Agency
- FDA
- Country
- United States
The hazard
Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
- Brand / company
- Siemens Healthcare Diagnostics Inc
- Codes / lots
- Material Number: 10329097. GTIN Number: 00630414482163. Lot Number: WW/08925. Expiration Date: 12/03/2025.
- Distribution
- worldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.
- Recall status
- Ongoing