The hazard

Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been reported. Use of the affected device may result hyperglycemia by failing to alert the user and delay treatment of diabetes.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

InPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users)

Brand / company
Medtronic MiniMed, Inc.
Codes / lots
UDI-DI: 0763000B000122075 CFN/Software Version: MMT-8060/7.0.0, 7.1.0, 7.2.0 and 7.5.0 MMT-8061/7.0.0, 7.0.1, 7.1.0, 7.1.1, 7.2.0 and 7.5.0
Distribution
International distribution to the countries of Argentina, Australia, Austria, Belgium, Bermuda, Chile, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Luxembourg, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 10 Sept 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.