PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20…
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane Product Codes and 12 more
- Published
- 1 Oct 2025
- Units affected
- 13 products
- Agency
- FDA
- Country
- United States
The hazard
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Allura Xper FD20/15; Product Code: 722058; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Allura Xper FD10/10 Product Codes: (1)722011, (2) 722027, (3) 722005; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Allura Xper FD20/15 OR Table; Product Code: 722059; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Allura Xper FD10 OR Table Product Codes: (1)722022, (2) 722033; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Allura CV20; Product Code: 722031; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Allura Xper FD10 Product Codes: (1)722010, (2) 722026, (3) 722003; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Allura Xper FD20 OR Table Product Codes: (1)722023, (2) 722035, (3) 722015; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Allura Xper FD20/20 OR Table; Product Code: 722039; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Allura Xper FD20 Biplane OR Table Product Codes: (1)722020, (2) 722025; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Allura Xper FD20 Product Codes: (1)722012, (2) 722028, (3) 722006; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Allura Xper FD20/20; Product Code: 722038; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
- Brand / company
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Codes / lots
- Product Codes: (1)722013, (2) 722008; UDI-DI: (1) 00884838059061, (2) N/A; Serial numbers: (1) 505, 313, 34, 153, 410, 360, 290, 241, 404, 126, 244, 106, 346, 383, 461, 307, 442, 435, 293, 289, 267, 214, 402, 86, 603, 149, 424, 140, 254, 255, 266, 518, 519, 303, 115, 150, 129, 496, 441, 124, 151, 550, 575, 562, 178, 330, 309, 601, 308, 264, 268, 141, 175, 85, 216, 422, 174, 228, 511, 152, 561, 262, 626, 497, 170, 379, 125, 564, 426, 495, 500, 515, 105, 428, 427, 531, 375, 316, 278, 347, 462, 249, 246, 639, 304, 147, 607, 567, 168, 171, 265, 332, 295, 589, 444, 447, 400, 384, 538, 117, 620, 624, 623, 215, 416, 415, 47, 335, 619, 598, 361, 358, 549, 133, 476, 315, 616, 630, 417, 490, 528, 525, 247, 618, 372, 613, 478, 544, 585, 608, 296, 570, 547, 138, 318, 345, 139, 473, 388, 434, 502, 586, 334, 411, 501, 458, 261, 429, 353, 395, 128, 348, 503, 557, 493, 563, 250, 595, 406, 355, 57, 394, 535, 569, 522, 465, 533, 412, 95, 344, 354, 425, 542, 207, 555, 524, 574, 366, 527, 325, 470, 440, 144, 403, 448, 134, 380, 487, 374, 256, 405, 367, 455, 302, 300, 299, 239, 506, 243, 324, 311, 70, 161, 449, 321, 176, 181, 554, 167, 130, 378, 158, 237, 453, 165, 532, 223, 343, 291, 370, 193, 163, 52
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 1 Oct 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.