PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10 and 14…
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10 and 14 more
- Published
- 8 Oct 2025
- Units affected
- 15 products
- Agency
- FDA
- Country
- United States
The hazard
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Allura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Allura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Allura Xper FD20/20 OR Table; Model Numbers: 722039; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Allura Xper FD10 OR Table; Model Numbers: 722033; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Allura Xper FD10C; Model Number: 722001; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Allura Xper FD20 Biplane OR Table; Model Numbers: (1) 722020, (2) 722025; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Allura Xper CV20; Model Numbers: 722031; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Allura Xper FD20 Biplane; Model Numbers: (1) 722008, (2) 722013; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Allura Xper FD20/20; Model Numbers: 722038; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Cardio Vascular-Allura Centron; Model Numbers: 722400; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Allura Xper FD20/15 OR Table; Model Numbers: 722059; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
- Brand / company
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Codes / lots
- Model Numbers: (1) 722003, (2) 722010, (3) 722026; UDI-DI: (1) N/A, (2) 00884838059030, (3) 00884838054189; Serial Numbers: (1) 1343, 1349, 1153, 1154, 1472, 257, 1694, 746, 745, 1599, 501, 503, 1360, 959, 1138, 490, 1188, 1532, 940, 1214, 401, 1627, 1031, 1728, 1189, 285, 1311, 1700, 998, 1187, 1222, 305, 1353, 939, 1326, 445, 1223, 700, 657, 527, 1634, 1023, 917, 1043, 994, 995, 945, 521, 590, 591, 855, 491, 1615, 1732, 1761, 1078, 367, 690, 1529, 385, 997616, 1085, 1663, 653, 1667, 928, 1449, 968, 1286, 72, 1622, 1621, 472, 1799, 463, 619915, 696, 1525, 1514, 574, 306, 1114, 1584, 447, 1950, 1647, 406, 1555, 1482, 1489, 1443, 1444, 1072, 1954, 1704, 1232, 749, 1266, 1089, 259, 1259, 1787, 345, 640, 37, 1714, 546, 1133, 1135, 1032, 1512, 892, 684, 857, 1084, 1067, 2936, 1411, 947, 197, 392, 1271, 1807, 1805, 1575, 963, 1430, 8198, 1869, 1594, 467, 1551, 1947, 390, 342, 368, 912, 913, 8091, 1471, 1474, 1446, 1001, 1092, 1718, 1637, 453, 833, 430, 1738, 1749, 1491, 1605, 1130, 5, 639, 674, 797, 1033, 1357, 1632, 1744, 281, 391, 1535, 350, 452, 629, 732, 754, 1038, 1045, 1536, 1056, 599, 1331, 756, 1351, 1611, 1610, 1804, 1122, 1413, 644, 545, 1059, 1327, 513, 1163, 1204, 1193, 1549
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 8 Oct 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.