NOXBOX LTD recalls NOxBOXi Nitric Oxide Delivery System
- Published
- 15 Oct 2025
- Units affected
- 1667 units
- Agency
- FDA
- Country
- United States
The hazard
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
- Brand / company
- NOXBOX LTD
- Codes / lots
- Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.
- Distribution
- Domestic: TN
- Recall status
- Ongoing