Sonesta Medical AB recalls Brand Name
- Published
- 19 Nov 2025
- Units affected
- 10 tables
- Agency
- FDA
- Country
- United States
The hazard
Video fluoroscopy table may contain inadequate weld on backrest bracket resulting in backrest dropping.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Brand Name: Sonesta Product Name: 6210 Model/Catalog Number: 0120-LA120-03 Software Version: N/A Product Description: Sonesta 6210 - Procedure table designed for full fluoroscopic imaging of the urinary tract during video Urodynamics procedures. Component: N/A
- Brand / company
- Sonesta Medical AB
- Codes / lots
- Lot Code: Primary DI Number: 07350006950018 Serial numbers of potentially affected units: 12-70791 12-70792 12-70793 12-70794 12-70795 12-70796 12-70797 12-70798 12-70799 12-70801
- Distribution
- US Distribution to NH only.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 19 Nov 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.