CheckMyRecalls

Tandem Diabetes Care, Inc. recalls Tandem Mobi Insulin Pump

Published
12 Nov 2025
Units affected
17,745
Agency
FDA
Country
United States

The hazard

Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). Pump, Tandem Mobi, Control-IQ/1010750, Pump, Tandem Mobi, Control-IQ, Replacement/1012719, Pump, Tandem Mobi, Control-IQ 7.6.0.3, Medicare/1013501, Pump, Tandem Mobi, Control-IQ 7.7/1013655, Pump, Tandem Mobi, Control-IQ, 7.7, Replacement/1013656, Pump, Tandem Mobi, Control-IQ 7.7, Medicare/1013700, Pharmacy Kit, Starter Pack, Tandem Mobi 7.7/1014081

Brand / company
Tandem Diabetes Care, Inc.
Codes / lots
Software versions: 7.6.0.1, 7.6.0.3, and 7.7.0.1 Catalog/UDI-DI: 1010750/00389152075013, 1012719/00389152271910, 1013501/00389152350110, 1013655/00389152365510, 1013656/00389152365619, 1013700/00389152370019, 1014081/00389152408170
Distribution
US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, PT, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 12 Nov 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.