Synthes (USA) Products LLC recalls LOW PROF LCKNG SCREW
- Published
- 29 Oct 2025
- Units affected
- 2 products
- Agency
- FDA
- Country
- United States
The hazard
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
- Brand / company
- Synthes (USA) Products LLC
- Codes / lots
- Product Code: 04.045.346TS. GTIN: 10886982301413. Lot Number: 31756P7. Expiration Date: 08/31/2025 | Product Code: 04.045.342TS. GTIN: 10886982301390. Lot Number: 31756P9. Expiration Date: 08/31/2025
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 29 Oct 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.