Philips Respironics, Inc. recalls SimplyGo Mini
- Published
- 19 Nov 2025
- Units affected
- 56 units (55 US, 1 OUS)
- Agency
- FDA
- Country
- United States
The hazard
Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600
- Brand / company
- Philips Respironics, Inc.
- Codes / lots
- Product Code: 1116816; UDI-DI: 00606959032385; Serial Numbers: 2438V115830, 2438V115960, 2438V116033, 2438V115899, 2438V115832, 2438V115964, 2438V116034, 2438V115837, 2438V115965, 2438V116040, 2438V115840, 2438V115967, 2438V116044, 2438V115842, 2438V115968, 2438V116045, 2438V115844, 2438V115970, 2438V116046, 2438V115878, 2438V115972, 2438V116047, 2438V115890, 2438V115973, 2438V116048, 2438V115894, 2438V115974, 2438V116049, 2438V115895, 2438V115980, 2438V116050, 2438V115896, 2438V115981, 2438V116051, 2438V115897, 2438V115983, 2438V116052, 2438V115919, 2438V115984, 2438V116226, 2438V115937, 2438V115991, 2438V116229, 2438V115939, 2438V115992, 2438V116233, 2438V115942, 2438V115993, 2438V116249, 2438V115948, 2438V115994, 2438V116253, 2438V115957, 2438V115995, 2438V116258, 2438V117762;
- Distribution
- US Nationwide distribution in the states of GA, ID, KY, NJ, NY and the country of Canada.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 19 Nov 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.