IMPRIS Imaging Inc recalls IMRIS Operating Suite
- Published
- 5 Nov 2025
- Units affected
- 4 products
- Agency
- FDA
- Country
- United States
The hazard
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
IMRIS Operating Suite, Neuro III-SV iMRI System, iMRX System, containing Siemens 3 Tesla MRI Magnet, Verio
IMRIS Operating Suite, InVision 3T Operating Suite, containing Siemens 3 Tesla MRI Magnet, Vida
IMRIS Operating Suite, IMRIS iMRI 3T V, containing Siemens 3 Tesla MRI Magnet, vIDA
IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra
- Brand / company
- IMPRIS Imaging Inc
- Codes / lots
- IMRIS Part Number(s): 105118-000, UDI/DI None predates UDI; 105118-001, UDI/DI None predates UDI; 105117-003, UDI/DI None predates UDI; 105117-004, UDI/DI None predates UDI; 105048-000, UDI/DI None predates UDI; 105048-001, UDI/DI None predates UDI; 105103-001, UDI/DI None predates UDI; 105071-004, UDI/DI None predates UDI; 105824-000, UDI/DI 00857534006189; 105828-000, UDI/DI 00857534006165; 116189-000, UDI/DI 00857534006202; 118396-000, UDI/DI 00857534006196. | IMRIS Part Number(s): 120823-000, UDI/DI 00857534006806 | IMRIS Part Number(s): 120823-000, UDI/DI 00857534006745 | IMRIS Part Number(s): 116053-000, UDI/DI 00857534006035; 116117-000, UDI/DI 00857534006042; 116118-000, UDI/DI 00857534006059; 116119-000, UDI/DI 00857534006066.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 5 Nov 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.