Inpeco S.A. recalls FlexLab (FLX)
- Published
- 3 Dec 2025
- Units affected
- 2 products
- Agency
- FDA
- Country
- United States
The hazard
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
FlexLab (FLX); Version: FLX-217-40;
FlexLab (FLX); Version: FLX-217-10;