CheckMyRecalls

Paragon 28, Inc. recalls Phantom Fibula Nail

Published
10 Dec 2025
Units affected
20
Agency
FDA
Country
United States

The hazard

Fibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S

Brand / company
Paragon 28, Inc.
Codes / lots
UDI-DI: 00889795133670. Lot: 501182207125. Expiration: March 13th, 2030
Distribution
US Nationwide distribution in the states of NJ, WA, NY, MD, CA, FL, NM, TX, IL, OH, NV, TN, LA.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 10 Dec 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.