Howmedica Osteonics Corp. recalls The Stryker CranialMask Tracker is a…

Howmedica Osteonics Corp. recalls The Stryker CranialMask Tracker is a single-use device consisting o…

Published
31 Dec 2025
Units affected
320 units
Agency
FDA
Country
United States

The hazard

When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.

Brand / company
Howmedica Osteonics Corp.
Codes / lots
Part Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896;
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 31 Dec 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.