AGFA Healthcare Corp. recalls Brand name
- Published
- 31 Dec 2025
- Units affected
- 35
- Agency
- FDA
- Country
- United States
The hazard
It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
- Brand / company
- AGFA Healthcare Corp.
- Codes / lots
- N/A
- Distribution
- U.S. Nationwide distribution.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 31 Dec 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.