CheckMyRecalls

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10F and 8…

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10F and 8 more

Published
21 Jan 2026
Units affected
16 products
Agency
FDA
Country
United States

The hazard

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Allura Xper FD10F; Model Number: 722002;

Allura Xper FD10/10; Model Number: 722027;

Allura Xper FD20; Model Number: 722028;

Product Name: Allura Xper FD10C; Model number: 722001;

Allura Xper FD10; Model Number: 722003;

Allura Xper FD10/10; Model Number: 722011;

Allura Xper FD20/20; Model Number: 722038;

Allura Xper FD10/10; Model Number: 722005;

Allura Xper FD20; Model Number: 722012;

Allura Xper FD20/10; Model Number: 722029;

Allura Xper FD10; Model Number: 722010;

Allura Xper FD20 Biplane; Model Number: 722008;

Brand / company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Codes / lots
Model Number: 722002; System Product Name: Allura Xper FD10F; UDI-DI: N/A; Serial Numbers: 61, 56, 27, 81, 25, 85, 20, 21, 84, 50, 31, 40; | Model Number: 722027; System Product Name: Allura Xper FD10/10; UDI-DI: 00884838054196; Serial Numbers: 298, 246, 146, 170, 238, 162, 7, 99, 81, 126, 55, 86, 6, 259, 205, 60, 57, 63, 145, 182, 48, 278; | Model Number: 722028; System Product Name: Allura Xper FD20; UDI-DI: 00884838054202; Serial Numbers: 893, 614; | Model Number: 722001; System Product Name: Allura Xper FD10C; UDI-DI: N/A; Serial Numbers: 166, 181, 153, 274, 296, 267, 206, 247, 161, 282, 213, 284, 103, 199, 129, 95, 299, 30, 235, 221, 348, 283, 276, 253, 1100, 78, 229, 232, 159, 177, 219, 217, 205, 3, 89, 101, 295, 305; | Model Number: 722003; System Product Name: Allura Xper FD10; UDI-DI: N/A; Serial Numbers: 1232, 967, 746, 1276, 232, 385, 1663, 1189, 81, 700, 167, 1072, 994, 150, 135, 1268, 168, 855, 213, 60, 471, 153, 319, 1159, 162, 180, 142, 1205, 719, 926, 426, 1413, 798, 65, 1165, 1658, 1193, 1778, 2933, 1585, 1380, 1706, 1613, 1685, 1297, 1388, 1522, 1238, 1416, 1448, 536, 1090, 1662, 1083, 1433, 1399, 62, 943, 249, 1227, 163, 1121, 1119, 889, 861, 1649, PP0200000118,
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 21 Jan 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.