ICU Medical, Inc. recalls ICU Medical Plum Duo Infusion Pump and 2 more
- Published
- 28 Jan 2026
- Units affected
- 3 products
- Agency
- FDA
- Country
- United States
The hazard
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)
ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)
ICU Medical Plum Duo Precision IV Pump, List Numbers: 400020403 (US), 400021003 (Philippines)
- Brand / company
- ICU Medical, Inc.
- Codes / lots
- 400020401 (US), Software Versions: 1.1.1, 1.1.2, 1.1.3, 1.1.4 400021001 (Philippines), Software Versions: 1.1.0 | 400010401 (US), Software Versions: 1.0.2; 400011001 (Philippines), Software Versions: 1.0.0 | 400020403 (US), Software Versions: 1.2.2; 400021003 (Philippines), Software Versions: 1.2.0
- Distribution
- Worldwide distribution - US Nationwide and the country of Philippines.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 28 Jan 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.