Maquet Cardiopulmonary Gmbh recalls Bubble Sensor for 3/8" x 3/32"…
Maquet Cardiopulmonary Gmbh recalls Bubble Sensor for 3/8" x 3/32" tubing
- Published
- 11 Feb 2026
- Units affected
- 3050 units
- Agency
- FDA
- Country
- United States
The hazard
Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
- Brand / company
- Maquet Cardiopulmonary Gmbh
- Codes / lots
- Model Number: 701055720; UDI-DI: 04037691816432; Serial Numbers: All units impacted;
- Distribution
- Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea, Republic of (South Korea), Latvia, Lithuania, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Uzbekistan, and Vietnam
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 11 Feb 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.