REPLIGEN CORPORATION recalls MAVERICK¿ Measurement Module Spectrometer…

REPLIGEN CORPORATION recalls MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally …

Published
21 Jan 2026
Units affected
33
Agency
FDA
Country
United States

The hazard

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Brand / company
REPLIGEN CORPORATION
Codes / lots
PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Distribution
U.S.
Recall status
Ongoing