Agiliti Health - Ellis recalls Adapt Pump
- Published
- 4 Mar 2026
- Units affected
- 4286
- Agency
- FDA
- Country
- United States
The hazard
Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure
- Brand / company
- Agiliti Health - Ellis
- Codes / lots
- UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025
- Distribution
- US Nationwide distribution.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 4 Mar 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.