Ion Beam Applications S.A. recalls IBA Proton Therapy System

Published
18 Mar 2026
Units affected
4 units (1 US, 3 OUS)
Agency
FDA
Country
United States

The hazard

It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

IBA Proton Therapy System - PROTEUS 235

Brand / company
Ion Beam Applications S.A.
Codes / lots
PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)
Distribution
Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
Recall status
Ongoing