CheckMyRecalls

Orthalign, Inc recalls Drill Plate

Published
1 Apr 2026
Units affected
2 products
Agency
FDA
Country
United States

The hazard

Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 1 -5 REF:402566 component Lantern Knee System

Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee System

Brand / company
Orthalign, Inc
Codes / lots
Lot # 25091103BB/ UDI: 00810832032427 | Lot # 25092204BB/ UDI:00810832032434
Distribution
US Nationwide distribution in the state of UT.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 1 Apr 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.