PHILIPS MEDICAL SYSTEMS recalls Philips Spectral CT on Rails
- Published
- 22 Apr 2026
- Units affected
- 3 systems
- Agency
- FDA
- Country
- United States
The hazard
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Philips Spectral CT on Rails. Model Number: 728334.
- Brand / company
- PHILIPS MEDICAL SYSTEMS
- Codes / lots
- Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.
- Distribution
- Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 22 Apr 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.