CheckMyRecalls

Heraeus Medical GmbH (Dental Division) recalls Heraeus

Published
6 May 2026
Units affected
2 products
Agency
FDA
Country
United States

The hazard

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Classification: Class III — this product is unlikely to cause health problems but breaks FDA rules (for example labeling).

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Heraeus, PALAMIX uno. Material Number: 66057893.

Heraeus, PALAMIX duo. Material Number: 66057897.

Brand / company
Heraeus Medical GmbH (Dental Division)
Codes / lots
Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060; | Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215;
Distribution
Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 6 May 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.