The hazard

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Philips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.

Philips SmartPath to Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 782118. 2. Model Number (REF): 782144. 3. Model Number (REF): 782163.

Philips Ingenia 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781342. 2. Model Number (REF): 781377. 3. Model Number (REF): 782103. 510(k) Numbers: K163116, K173079, K183063, K193215, K213516, K213583, K223458, K251397, K251808

Philips Evolution upgrade 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782143. 2. Model Number (REF): 782162.

Philips Ingenia Ambition S with MR Elastography (MRE). 1. Model Number (REF): 781359. 2. Model Number (REF): 782108. 3. Model Number (REF): 782139.

Philips Achieva 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781196. 2. Model Number (REF): 781296.

Philips Ingenia 1.5T CX with MR Elastography (MRE). 1. Model Number (REF): 781262.

Philips Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 781358. 2. Model Number (REF): 782107. 3. Model Number (REF): 782136. 510(k) Numbers: K213516, K223458, K230972, K232030, K251397, K251808

Philips Ingenia 3.0T CX with MR Elastography (MRE). 1. Model Number (REF): 781271. 2. Model Number (REF): 782105. 510(k) Numbers: K173079, K183063, K193215, K213583, K223458, K251397, K251808.

Philips Achieva 1.5T Initial system with MR Elastography (MRE). 1. Model Number (REF): 781178.

Philips Ingenia 1.5T with MR Elastography (MRE). 1. Model Number (REF): 781315. 2. Model Number (REF): 781341. 3. Model Number (REF): 781396. 4. Model Number (REF): 782115. 510(k) Numbers: K163116, K173079, K183063, K193215, K213583, K223458, K251397, K251808

Philips MR 7700 with MR Elastography (MRE). 1. Model Number (REF): 782120. 2. Model Number (REF): 782153.

Brand / company
Philips North America
Codes / lots
1. Model Number (REF): 782145. UDI: (01)00884838108684(21)17385. Serial Numbers: 17385. | 1. Model Number (REF): 782118. UDI: (01)00884838099746(21)85655. Serial Numbers: 85655. 2. Model Number (REF): 782144. UDI: (01)00884838108677(21)71087. Serial Numbers: 71087. 3. Model Number (REF): 782163. UDI: (01)00884838115743(21)42299. Serial Numbers: 42299. | 1. Model Number (REF): 781342. UDI: (01)00884838055339(21)71693, (01)00884838055339(21)71581, (01)00884838055339(21)71582, (01)00884838055339(21)71607, (01)00884838055339(21)85216, (01)00884838055339(21)71106, (01)00884838055339(21)71477, (01)00884838055339(21)71512, (01)00884838055339(21)71966, (01)00884838055339(21)71967, (01)00884838055339(21)85147, (01)00884838055339(21)85224, (01)00884838055339(21)85318, (01)00884838055339(21)71462, (01)00884838055339(21)71619, (01)00884838055339(21)85250, (01)00884838055339(21)71002, (01)00884838055339(21)71037, (01)00884838055339(21)71269, (01)00884838055339(21)71841, (01)00884838055339(21)71885, (01)00884838055339(21)71897, (01)00884838055339(21)85126, (01)00884838055339(21)85109, (01)00884838055339(21)71296, (01)00884838055339(21)71188, (01)00884838055339(21)71088, (01)00884838055339(21)713
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.
Recall status
Ongoing