PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion systems…

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion systems not configured with an optional auxiliary p…

Published
27 May 2026
Units affected
5,537 units
Agency
FDA
Country
United States

The hazard

Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descriptions and corresponding model numbers. 1. Azurion 3 M12; Model Numbers: 722063, 722229. 2. Azurion 3 M15; Model Numbers: 722222, 722064, 722280, 722230. 3. Azurion 5 M12; Model Numbers: 722227, 722231. 4. Azurion 5 M20; Model Numbers: 722228, 722232, 722281. 5. Azurion 7 B12; Model Numbers: 722067, 722225, 722235. 6. Azurion 7 B20; Model Numbers: 722068, 722226, 722236. 7. Azurion 7 M12; Model Numbers: 722078, 722223, 722233. 8. Azurion 7 M20; Model Numbers: 722079, 722224, 722234, 722282.

Brand / company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Codes / lots
1. Model Number [UDI]: 722063 [(01)00884838085275(21)77, (01)00884838085275(21)47, (01)00884838085275(21)146, (01)00884838085275(21)102, (01)00884838085275(21)43, (01)00884838085275(21)2, (01)00884838085275(21)99, (01)00884838085275(21)28, (01)00884838085275(21)144, (01)00884838085275(21)62, (01)00884838085275(21)119, (01)00884838085275(21)61, (01)00884838085275(21)29, (01)00884838085275(21)40, (01)00884838085275(21)70, (01)00884838085275(21)120, (01)00884838085275(21)121, (01)00884838085275(21)110, (01)00884838085275(21)53, (01)00884838085275(21)103, (01)00884838085275(21)97, (01)00884838085275(21)101, (01)00884838085275(21)143, (01)00884838085275(21)116, (01)00884838085275(21)93, (01)00884838085275(21)141, (01)00884838085275(21)115, (01)00884838085275(21)137, (01)00884838085275(21)150, (01)00884838085275(21)128, (01)00884838085275(21)4, (01)00884838085275(21)50, (01)00884838085275(21)129, (01)00884838085275(21)78, (01)00884838085275(21)76, (01)00884838085275(21)69, (01)00884838085275(21)59, (01)00884838085275(21)92, (01)00884838085275(21)139, (01)00884838085275(21)130, (01)00884838085275(21)154, (01)00884838085275(21)75, (01)00884838085275(21)24, (01)00884838085275(21)94, (01)008
Distribution
Nationwide distribution. International distribution to Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Chile, C¿te D'Ivoire, El Salvador, Estonia, Greece, Honduras, Hong Kong, Iceland, Indonesia, Iraq, Japan, Kyrgyzstan, Malaysia, Malta, Moldova, Nepal, New Zealand, Philippines, Qatar, Singapore, South Africa, Sri Lanka, Syrian Arab Republic, Tanzania, Turkmenistan, Vietnam, Yemen
Recall status
Ongoing