Fresenius Kabi USA, LLC recalls LVP software of the Ivenix Infusion…
Fresenius Kabi USA, LLC recalls LVP software of the Ivenix Infusion System (IIS)
- Published
- 10 Jun 2026
- Units affected
- 32 systems
- Agency
- FDA
- Country
- United States
The hazard
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
- Brand / company
- Fresenius Kabi USA, LLC
- Codes / lots
- Product Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2
- Distribution
- US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
- Recall status
- Ongoing